BOSTON SCIENTIFIC CORPORATION DIREXION HI-FLO; CATHETER, CONTINUOUS FLUSH Back to Search Results
Model Number 80619
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/25/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported nitinol coating came off.The target lesion was located in a vessel in the uterus area.A 155cm direxion hi-flo was selected for use.During the procedure, it was noted that the nitinol coating came off at about 20cm from proximal yellow hub.Subsequently, the physician lost all tracking but was able to complete the procedure with the device.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by manufacturer: the device was returned for analysis.The device was inspected for any damage or irregularities.The direxion had a fracture 1 mm from the strain relief and kinks 10.6 cm and 106.4 cm from the strain relief.The strain relief was not removed from the hub due to the location of the fracture.The shaft was not completely separated, the inner liner was still intact.The fracture looked consistent with a bending and tensile force breaking.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.
 
Event or Problem Description
It was reported nitinol coating came off.The target lesion was located in a vessel in the uterus area.A 155cm direxion hi-flo was selected for use.During the procedure, it was noted that the nitinol coating came off at about 20cm from proximal yellow hub.Subsequently, the physician lost all tracking but was able to complete the procedure with the device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIREXION HI-FLO
Common Device Name
CATHETER, CONTINUOUS FLUSH
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
model farm road
cork
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key14168701
Report Number2134265-2022-04221
Device Sequence Number6028484
Product Code KRA
UDI-Device Identifier08714729839491
UDI-Public08714729839491
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K142259
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/27/2024
Device Model Number80619
Device Catalogue Number80619
Device Lot Number0028756877
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/12/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/25/2022
Supplement Date Received by Manufacturer05/06/2022
Initial Report FDA Received Date04/21/2022
Supplement Report FDA Received Date05/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/27/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1