It was reported that during a thoracoscopic left lobectomy, the sleeve was broken off while a camera was operated after the device was used for about 2 hours.The device was used on the mid-axillary line in the 6th.The device was removed from the patient and checked, then it was found that a 1mm fragment was left inside the patient.The intrathoracic lavage was performed but it could not be found.The surgeon thought it is not a problem because it was very small.The procedure was not converted.The patient¿s condition is stable.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
|
(b)(4).Date sent: 6/6/2022 investigation summary the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that the tt012 device was received with the tip of the sleeve cracked.The event reported was confirmed and it is related to improper use of the device.One possible cause for the damage found on the sleeve may be excessive external load placed on the device.Please reference the instruction for use for more information.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Additional information was requested, and the following was obtained: can you describe the size and shape of the piece that was broken off? => about 1.5mm what type of instruments were exchanged through the trocar sleeve? => when the trocar sleeve broken, the camera was used.Were any sharp instruments exchanged through the trocar sleeve? => no further information is available.Are pictures of the device available? =>the device was returned.Was there any alleged deficiency of trocar sleeve? => no.What is the current patient status? =>the patient's condition is stable.No further information will be provided.
|