| Model Number |
JB725 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a patient underwent an unknown procedure on (b)(6) 2022 and suture was used.During the procedure, the needle was detached from the suture easily during interrupted suturing on fascia.This happened many times.It was a control release needle.There were no patient consequences reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The single complaint was reported with multiple events.There are no additional details regarding the additional events.Additional information was requested, and the following was obtained: was there any patient consequences? could you please advise how many times the needle was detached from the suture? no further information is available.Did the event occur during one or multiple patient procedures? during one procedure.What is the total number of procedures? one.Please provide the lot number: no further information is available.Device return status: we regularly contact with sales rep about the device returning.No further information will be provided.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).Date sent to the fda: 06/20/2022.H6 component code: g07002 - no device problem found.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: h6, d9, h3.H3 investigation summary: one opened box and five packets of product code jb725 were returned for analysis with the packaging closed.Upon initial inspection, of the samples, no external damages were observed on the packet.The product code is a control release and requires eight strands per packet.In order to evaluate the conditions of the returned samples, the five packets were opened, and no defects were detected.The swage and attachment area was noted to be as expected.The sutures were dispensed without problems and examined along the strand to detect any issues related to no defects observed during evaluation.Functional test was performed, and the pull force result meets the customer requirements.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.As part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.There may have been other circumstances or issues that occurred during the use of the device that we were unable to duplicate during our laboratory analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|