(b)(4).Batch #: unk.A manufacturing record evaluation was performed for the finished device lot number, and no [non-conformances/ manufacturing irregularities] were identified.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
(b)(4) date sent: 6/15/2022 d4: batch # 337a25 investigation summary the product was returned to ethicon endo-surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the ats45nk device was returned with no apparent damage and with no reload present on the device.The device was tested for functionality with a test reload, it fired and formed the staples as intended.The staple line was complete and the staples meet the staple form release criteria.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: when positioning the stapler on the application site, ensure that no obstructions such as clips, stents, guidewires, etc., are within the instrument jaws.Firing over an obstruction may result in incomplete cutting action, improperly formed staples, and/or inability to open the instrument jaws.Incomplete firing may result in malformed staples, bleeding, and/or difficulty removing the device.The event described could not be confirmed as the device performed without any difficulties noted and no cartridge was returned for analysis.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|