WRIGHTS LANE SYNTHES USA PRODUCTS LLC 3.5 VA LCKNG SCR SLF-TPNG/SD/42; SCREW, FIXATION, BONE Back to Search Results
Model Number 02.127.142
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/24/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #(b)(4).If the information is unknown, not available or does not apply, the section/field of the form is left blank.Complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that while attempting to insert a 3.5 va locking screw into a va locking hole, the screw would not lock into the plate.A 3.5 cortex screw was placed in the hole and the surgery was completed successfully without any surgical delay.Additionally, it was reported that a periarticular reduction forceps was difficult to open.However, it opened freely once it was wet with irrigation.It was believed to have bioburden.The forceps remained stiff postoperatively after washing.This report is for the locking screw.This is report 1 of 2 for complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3.5 VA LCKNG SCR SLF-TPNG/SD/42
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
WERK GRENCHEN (CH)
solothurnstrasse 186
grenchen 2540
SZ   2540
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key14171775
Report Number2939274-2022-01442
Device Sequence Number12197988
Product Code HWC
UDI-Device Identifier10886982043252
UDI-Public10886982043252
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K121601
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number02.127.142
Device Catalogue Number02.127.142
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/24/2022
Initial Report FDA Received Date04/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
3.5 VA PROX TIB PL SMALL BEND/4H/87/LFT; PERIARTIC-REDUCT-FORCEPS MEDIUM; UNK - SCREWS: 3.5 MM CORTEX