Model Number QC-3-8-3D
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/19/2022
Type of Reportable Event Malfunction
Event or Problem Description
Medtronic received information reporting that prior to use in the procedure, the tip of the axium coil was found to be damaged.It was noted the coil was prepared according to the instructions for use (ifu).The coil was not used and was replaced with another manufacturer's device for the procedure.As the issue was observed prior to use, there was no patient involvement.Additional information was received indicating there were no issues that resulted in the damage.
 
Additional Manufacturer Narrative
Product analysis: equipment used: video inspection system (m-78210), ruler 200cm (m-83361), camera (panasonic lumix dmc-zs5) as found condition: the axium coil was returned for analysis within ashipping box; and within a plastic bio-pouch.Visual inspection/damage location details: the axium coil was returned without its introducer sheath.Therefore, any contributing factors from the introducer sheath could not be assessed.The axium pushwire was found to be broken at ~3.5cm from proximal end with release wire pulled.The axium implant coil appeared to be detached from pushwire.In addition, the axium coil pulled back from lumen stop and shield coil stretched.No other anomalies were observed.Conclusion: based on the device analysis and reported information, the customer¿s report of ¿coil kink/damage¿ was confirmed.In addition, the axium coil was also found to be detached from the pushwire.However, the root cause could not be determined.The review of lot history records shows that the finished device has met all manufacturing requirements and specifications during final assembly and quality inspection.There was no non-conformance to specification identified that led to the coil kink/damage issue.The event is reported at this time based on analysis results which indicated a reportable event.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIUM 3D
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key14172148
Report Number2029214-2022-00683
Device Sequence Number12233092
Product Code KRD
UDI-Device Identifier00847536029897
UDI-Public00847536029897
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K081465
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/07/2023
Device Model NumberQC-3-8-3D
Device Catalogue NumberQC-3-8-3D
Device Lot NumberB050654
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/28/2022
Initial Report FDA Received Date04/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/07/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-