• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. BALLARD CS NEO/PED, 8 FR, ELBOW; VAP CLOSED SUCTION CATHETERS & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. BALLARD CS NEO/PED, 8 FR, ELBOW; VAP CLOSED SUCTION CATHETERS & ACCESSORIES Back to Search Results
Model Number 208
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code Decreased Respiratory Rate (2485)
Date of Event 03/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The actual complaint product was not returned for evaluation.A review of the device history record is in-progress.All information reasonably known as of 20-apr-2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).
 
Event or Problem Description
It was reported the etco2 [end-tidal carbon dioxide] was reading low as well as exhaled volume.The patient's breath sounds were clear and bilaterally with no audible leak heard.The patent's trach cuff was deflated and reinflated.Display on etco2 showed a much lower rr [respiratory rate] than the patient's actual rr so all end tidal disposable equipment was replaced, and a new module was used with no change in etco2 or rr reading.No audible leak was noted, and the in-line suction catheter was able to pass and functioned.The patient was much less arousable than usual.An i-stat [blood analysis test] was performed using venous blood showing normal values.The ventilator was then changed to a v500 machine and problem of high leak persisted.The tracheostomy was then changed to rule out airway dislodgement to no avail.The only part of the patient ventilator circuit that was not replaced yet was the in-line suction catheter and after doing that it was noticed that the suction button did not fully spring back to completely off leaving a continuous suction to the airway.
 
Additional Manufacturer Narrative
The device history record for the reported lot number, 30150009, was reviewed and documented that the lot was manufactured according to the approved manufacturing procedures, specifications, and then released by quality assurance.All information reasonably known as of 02-feb-2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BALLARD CS NEO/PED, 8 FR, ELBOW
Common Device Name
VAP CLOSED SUCTION CATHETERS & ACCESSORIES
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT S DE RL DE CV
carretera internacional
salida norte no. 1053
magdalena, cp
Manufacturer Contact
lisa clark
5405 windward parkway
alpharetta, GA 30004
4704485444
MDR Report Key14173041
Report Number8030647-2022-00129
Device Sequence Number20086335
Product Code BSY
UDI-Device Identifier00609038938332
UDI-Public00609038938332
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/04/2026
Device Model Number208
Device Catalogue NumberN/A
Device Lot Number30150009
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/23/2022
Supplement Date Received by Manufacturer01/05/2023
Initial Report FDA Received Date04/21/2022
Supplement Report FDA Received Date02/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
V500 VENTILATOR MACHINE.
Patient Age3 YR
Patient SexMale
Patient Weight14 KG
-
-