The actual complaint product was not returned for evaluation.A review of the device history record is in-progress.All information reasonably known as of 20-apr-2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).
|
|
It was reported the etco2 [end-tidal carbon dioxide] was reading low as well as exhaled volume.The patient's breath sounds were clear and bilaterally with no audible leak heard.The patent's trach cuff was deflated and reinflated.Display on etco2 showed a much lower rr [respiratory rate] than the patient's actual rr so all end tidal disposable equipment was replaced, and a new module was used with no change in etco2 or rr reading.No audible leak was noted, and the in-line suction catheter was able to pass and functioned.The patient was much less arousable than usual.An i-stat [blood analysis test] was performed using venous blood showing normal values.The ventilator was then changed to a v500 machine and problem of high leak persisted.The tracheostomy was then changed to rule out airway dislodgement to no avail.The only part of the patient ventilator circuit that was not replaced yet was the in-line suction catheter and after doing that it was noticed that the suction button did not fully spring back to completely off leaving a continuous suction to the airway.
|
|
The device history record for the reported lot number, 30150009, was reviewed and documented that the lot was manufactured according to the approved manufacturing procedures, specifications, and then released by quality assurance.All information reasonably known as of 02-feb-2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
|