| Model Number |
STEM EXTENSION 22MM X 80 L. |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
02/08/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
As reported, approximately 13.5 years post tka, this 80 y/o male patient experienced increasing pain and left leg no longer able to bear weight.Radiologically showed periprosthetic fracture with fracture of the femoral stem at the adapter/stem transition.Patient had a revision and the explant was preserved.No additional information available.
|
| |
|
Additional Manufacturer Narrative
|
|
(h3) the revision reported may have been the result of femoral loosening overtime, which created a bending moment on the femoral stem extension and led to crack initiation, propagation, and ultimate fracture at the threaded junction of the femoral stem extension and a periprosthetic fracture.However, this cannot be confirmed as the devices were not returned for evaluation and photographs and x-rays were unable to be obtained.The most probable root cause associated with the reported event of "prosthesis fracture¿ is associated with weakened integration of the device at the bone-implant interface due to loss of fibrous and/or bony tissue and leading to compromised anchorage of the device.
|
| |
|