• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (SPARKS) BD BBL¿ MUELLER HINTON II AGAR; CULTURE MEDIA, ANTIMICROBIAL SUSCEPTIBILITY TEST, MUELLER HINTON AGAR/BROTH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (SPARKS) BD BBL¿ MUELLER HINTON II AGAR; CULTURE MEDIA, ANTIMICROBIAL SUSCEPTIBILITY TEST, MUELLER HINTON AGAR/BROTH Back to Search Results
Model Number 221800
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that while using bd bbl¿ mueller hinton ii agar the customer observed that the media had atypical growth.This event occurred 1 time(s).There is no indication that false results were reported to health care practitioners.There is no indication of patient impact.The following information was provided by the initial reporter: "[customer] reported further growth issues with their m.Luteus strain when testing the new lot of mueller hinton ii agar plates received (lot 2020098 / 0800/ 2022/05/02).According to the attached photo, their m.Luteus strain is growing well on their old (expired) lot 1211750 but it's not growing on the two last lots received (1347273 and 2020098).The same broth tube was used to inoculate the 3 lots and the plates were all incubated at 37 degrees c for 48h.".
 
Event or Problem Description
It was reported that while using bd bbl¿ mueller hinton ii agar the customer observed that the media had atypical growth.This event occurred 1 time(s).There is no indication that false results were reported to health care practitioners.There is no indication of patient impact.The following information was provided by the initial reporter: "[customer] reported further growth issues with their m.Luteus strain when testing the new lot of mueller hinton ii agar plates received (lot 2020098 / 0800/ 2022/05/02).According to the attached photo, their m.Luteus strain is growing well on their old (expired) lot 1211750 but it's not growing on the two last lots received (1347273 and 2020098).The same broth tube was used to inoculate the 3 lots and the plates were all incubated at 37 degrees c for 48h.".
 
Additional Manufacturer Narrative
H.6 investigation summary: during manufacturing of material 221800, media is formulated and sent through a high temperature short time sterilizer to remove bioburden.The petri dishes are subjected to uv radiation to decrease bioburden.The petri dishes are filled in a positive pressure hepa filtered environment.The filled plates are cooled and immediately wrapped into sleeves to decrease the introduction of microbes.Sleeves are then packaged into cartons and then transferred to a refrigerated truck (2 to 8 degrees c) for shipment to the distributor.Bd distributors are provided with the storage guidelines for the shipping and handling of bd media of 2 to 8 degrees c in a dark place.The batch history record for batch 2020098 was satisfactory and no quality notifications were generated during manufacturing and inspection.All batches are tested prior to release and results reported on the certificate of analysis which can be obtained at www.Bd.Com/regdocs.Mueller hinton agar is stability tested biennially for biological performance to ensure satisfactory performance throughout shelf life with the organisms that are reported on the certificate of analysis.All performance testing on this batch was satisfactory at the time of release.The complaint history was reviewed, and no other complaints have been taken on batch 2020098.Retention samples from batch 2020098 were not available for inspection.One photo was received for investigation.The photo shows the bottom of six plates, each with eight discs stamped onto the surface but only two plates have visible growth.No plate prints can be read for batch verification from the photos.It is followed that there was no growth where expected however, conclusions about performance cannot be made from photos.No samples were received for investigation.This complaint cannot be confirmed.Bd will continue to trend complaints for performance.Risk management file review assessed the potential risk for the defect as severity s3 per baltrmppmgenpuraph, rev 02, id 1.6.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD BBL¿ MUELLER HINTON II AGAR
Common Device Name
CULTURE MEDIA, ANTIMICROBIAL SUSCEPTIBILITY TEST, MUELLER HINTON AGAR/BROTH
Manufacturer (Section D)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer (Section G)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14173966
Report Number1119779-2022-00584
Device Sequence Number17377431
Product Code JTZ
UDI-Device Identifier10382902218002
UDI-Public10382902218002
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date05/02/2022
Device Model Number221800
Device Catalogue Number221800
Device Lot Number2020098
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/28/2022
Supplement Date Received by Manufacturer10/26/2022
Initial Report FDA Received Date04/21/2022
Supplement Report FDA Received Date11/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/20/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-