BECTON DICKINSON MEDICAL SYSTEMS 3 ML BD INTEGRA¿ RETRACTING SAFETY SYRINGE WITH 21 G X 1 1/2 IN.; PISTON SYRINGE Back to Search Results
Model Number 305274
Medical Device Problem Code Retraction Problem (1536)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/21/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that while using a 3 ml bd integra¿ retracting safety syringe with 21 g x 1 1/2 in., the needle failed to retract.The following information was provided by the initial reporter: syringe used did not retract from muscle, although the mechanisms clicked as though to retract.
 
Additional Manufacturer Narrative
Initial reporter phone#: unknown.Initial reporter fax#: unknown.Investigation summary: since no samples displaying the condition reported are available for examination, we were unable to fully investigate this incident.No root cause can be determined as no samples were received.Furthermore, a device history record review showed no rejected inspections or quality issues during the production of the provided lot number that could have contributed to the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3 ML BD INTEGRA¿ RETRACTING SAFETY SYRINGE WITH 21 G X 1 1/2 IN.
Common Device Name
PISTON SYRINGE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer (Section G)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14176678
Report Number1213809-2022-00228
Device Sequence Number14526457
Product Code MEG
UDI-Device Identifier30382903052746
UDI-Public30382903052746
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K011103
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date01/31/2026
Device Model Number305274
Device Catalogue Number305274
Device Lot Number1029961
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/31/2022
Initial Report FDA Received Date04/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-