• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (SPARKS) BD BACTEC¿ FX, INSTRUMENT BOTTOM, PACKAGED; SYSTEM, BLOOD CULTURING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (SPARKS) BD BACTEC¿ FX, INSTRUMENT BOTTOM, PACKAGED; SYSTEM, BLOOD CULTURING Back to Search Results
Model Number 441386
Medical Device Problem Code False Positive Result (1227)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/31/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medical device expiration date: na.A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.
 
Event or Problem Description
It has been reported that the bd bactec¿ fx, instrument bottom, packaged has been found producing a false positive result.There was no report of confirmatory testing.There was no report of patient impact.The following has been provided by the initial reporter: customer is reporting 4-6 anaerobic 442021 bottles that are false positive.
 
Event or Problem Description
It has been reported that the bd bactec¿ fx, instrument bottom, packaged has been found producing a false positive result.There was no report of confirmatory testing.There was no report of patient impact.The following has been provided by the initial reporter: customer is reporting 4-6 anaerobic 442021 bottles that are false positive.
 
Additional Manufacturer Narrative
H.6 investigation summary: a complaint of "4 - 6 442021 bottles false positive detected" was received against instrument bottom bactec fx packaged material number: 441386, serial number: (b)(6).Instrument was found to be functional, and the customer started using the instrument for regular operation.This is an unconfirmed complaint, and the root cause was unable to be determined.Device history record review for the instrument (b)(6), is not required for this complaint as this complaint does not allege an early life failure or failure at installation and the configuration has changed since release from manufacturing due to service repairs/pms.Service history review revealed no previous complaint for this issue.No parts were replaced as part of this complaint, and therefore no samples were returned, and no returned material investigation could occur.Complaints for "results" are under statistical control for the month of april 2022.The upper control limit was not breached.Quality will continue to monitor the ¿results".No new trends have been identified.Review of risk management documentation indicates that the potential risk of the reported failure mode was assessed as severity s1 (user inconvenience) via document baltrmbactecinstraph revision 15 id 1.2c.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD BACTEC¿ FX, INSTRUMENT BOTTOM, PACKAGED
Common Device Name
SYSTEM, BLOOD CULTURING
Manufacturer (Section D)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer (Section G)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14178011
Report Number1119779-2022-00588
Device Sequence Number5783337
Product Code MDB
UDI-Device Identifier00382904413866
UDI-Public00382904413866
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K915796
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number441386
Device Catalogue Number441386
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/31/2022
Supplement Date Received by Manufacturer11/03/2022
Initial Report FDA Received Date04/21/2022
Supplement Report FDA Received Date11/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/13/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-