Model Number MICL 13.2
Medical Device Problem Codes Inadequacy of Device Shape and/or Size (1583); Unintended Movement (3026)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 03/21/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Claim#: (b)(4).
 
Event or Problem Description
The reporter indicated that a 13.2mm micl 13.2 implantable collamer lens of -16.0 diopter was implanted into the patients left eye (os) on (b)(6) 2022.On (b)(4) 2022 the lens was removed due to low vault.Upon removal of the initial lens it was determined to flipped inside the eye.Reporter states there was difficulty tucking haptics under the iris as the manipulator constantly stuck to the icl.On the same day different surgery a replacement lens of same model/size and diopter was implanted and the problem was resolved.Status of the eye reported as moderate anterior cell.Mild corneal edema.Patent pi.Per reporter "it is typical for any post operative eye to have some cell and corneal edema, especially after being operated on a second time.The patent pi is also normal since we did not have an evo lens so we had to do the pi.".
 
Additional Manufacturer Narrative
H6- method code 3331: device history record (dhr) review: based on the results of the investigation, all released devices from the associated work orders(s), including the suspected device, have been manufactured within established parameters; and there is no indication that the manufacturing and processing of the device contributed to the complaint issue.Claim# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene enriquez
1911 walker avenue
monrovia, CA 91016
MDR Report Key14178237
Report Number2023826-2022-01223
Device Sequence Number5025434
Product Code MTA
UDI-Device Identifier00841542103411
UDI-Public00841542103411
Combination Product (Y/N)N
Initial Reporter StateDC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2023
Device Model NumberMICL 13.2
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/22/2022
Supplement Date Received by Manufacturer04/25/2022
Initial Report FDA Received Date04/21/2022
Supplement Report FDA Received Date06/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/03/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL: SFC-45: LOT# UNK.; FOAM TIP PLUNGER (FTP), LOT# UNK.; INJECTOR MODEL: MSI-PF, LOT# UNK.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age34 YR
Patient SexFemale
-
-