| Model Number |
1804275-18 |
| Medical Device Problem Codes |
Material Separation (1562); Deformation Due to Compressive Stress (2889)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/29/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a lesion located in the diagonal artery that was moderately calcified.The xience skypoint shaft kinked and broke during insertion into 6fr guide catheter.There was no reported resistance during advancement and the device was easily removed.There was no reported adverse patient effect and no clinically significant delay in the procedure.Another xience skypoint was used to complete the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
| |
|