| Model Number |
ASC4630-01 |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/15/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, during an arthroscopic anterior cruciate ligament reconstruction, the head of the tristar blade was deformed during the operation, and the metal dots fell off.The pieces were removed from the patient using forceps.Surgery was resumed, with a back-up device, after a non-significant delay.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device was received for evaluation.A visual inspection of the returned instrument shows no manufacturing abnormalities.The electrode has one metal ball wire detached.Product was out of the original packaging.No packaging returned.The device was plugged into the controller and registered settings.The wand was able to generate plasma and coagulation as intended with the remaining electrode.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A clinical review states per case details, the broken pieces were retrieved from the patient.The procedure was completed using a back-up device.No patient injuries or adverse consequences were reported.Since no patient injuries are being reported no further clinical/medical assessment is warranted at this time.It was determined the device contributed to the reported event.The complaint was confirmed and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include: (1) hitting the device tip against a hard surface.(2) using the device as a lever to enlarge surgical site.(3) mechanical displacement of tissue through applied force.No containment or corrective actions are recommended at this time.
|
| |
|
Search Alerts/Recalls
|