| Model Number |
9013S0042 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial report# (b)(4).On several different patients, during voluntary contraction (mainly iliopsoas, quadriceps and medial gastrocnemius muscles), the needles used were twisted).No injuries reported.Further investigation to be carried out.
|
| |
|
Event or Problem Description
|
|
On several different patients, during voluntary contraction (mainly iliopsoas, quadriceps and medial gastrocnemius muscles), the needles used were twisted.No injuries reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Follow up report 001 #10153070.Pictures were provided from the customer of the affected parts.Affected parts to be returned from the customer.Parts to be evaluated and investigation results to be documented once they become available.
|
| |
|
Event or Problem Description
|
|
On several different patients, during voluntary contraction (mainly iliopsoas, quadriceps and medial gastrocnemius muscles), the needles used were twisted.No injuries reported.
|
| |
|
Event or Problem Description
|
|
On several different patients, during voluntary contraction (mainly iliopsoas, quadriceps and medial gastrocnemius muscles), the needles used were twisted.No injuries reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Follow up report 002 reference natus complaint #(b)(4).Returns were received from the customer and visual inspection was carried out by engineering.No issues noted during inspection.It was confirmed that needles did not have blunt tips.There are no capa's related to this issue.The current risk file (b)(6) rev l hazard id 6.5 identifies this issue harm - patient discomfort during procedure may increase.Cause- needle is perceived as being more difficult to insert than expected / increased force required to insert the needle due to high friction severity- marginal (3) risk level- minor (3) a risk review is not required as this complaint does not describe a new failure mode or new harm and the existing hazard severity and/or probability of occurrence has not changed.Complaints are reviewed routinely per quality system requirements (b)(4) corporate trending and analysis procedure) complaint trends are assessed as part of these reviews.A review of complaint trending is completed quarterly.Last review (b)(6)rend for this failure identified.Device history record - wo#322377 was reviewed.No issues noted during production.Install date of device release date: (b)(6) - 20/10/2021 a search for service data that may be relevant to this issue has been conducted.No further information related to this issue found.Closure rationale: complaint could not be verified, monitor for future occurrence.
|
| |
|
Search Alerts/Recalls
|