Model Number 013562
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/30/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during an arthroscopy procedure using a bone tunnel plug, when the femoral and tibial tunnel were made, the doctor decided to relocate the femoral pin and therefore used a 10mm tibial plug.At the time of removing the plug, it fractured inside the joint.It was mentioned that no excessive force was used when removing the plug, it was done by hand.An attempt was made to remove the fragment of the plug that was left interarticular with forceps, and it broke again into multiple small parts.An exhaustive washing of the joint was performed and all the fragments were removed.It is unknown how the procedure was completed.A surgical delay of 30 min was reported.No other complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).H6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.There was no way to determine if the device contributed to the reported event.A complaint history review found no similar reported events.Insufficient product identification information was provided, and thus, a manufacturing record review could not be conducted.Based on the information available, there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Based on the information provided, an exhaustive washing of the joint was performed, and it was reported all the fragments were removed from inside of the patient.According to the report, there was a surgical delay of 30 minutes, however, it is unknown how the procedure was completed.Since there were no other complications were reported, no further clinical/medical assessment is warranted at this time.Should any additional relevant medical investigation be provided, this case would be re-assessed.The complaint was not confirmed.Factors that could have contributed to the reported event include an application of unintended inappropriate or excessive force to the device, an impact event inconsistent with normal use, wear from prolonged use, misuse or rough handling, or inadequate routine maintenance.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BONE TUNNEL PLUG LARGE 10-11MM
Common Device Name
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin 78735
5123913905
MDR Report Key14184288
Report Number1219602-2022-00585
Device Sequence Number12018093
Product Code NBH
UDI-Device Identifier03596010091147
UDI-Public03596010091147
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number013562
Device Catalogue Number013562
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/31/2022
Supplement Date Received by Manufacturer05/24/2022
Initial Report FDA Received Date04/22/2022
Supplement Report FDA Received Date05/25/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-