Model Number VERSION 1.0 |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/31/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
As the lot number for the device was provided, a review of the device history record is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.The medical device manufacturer and manufacturing location for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.(expiry date: 09/2023).
|
|
Event Description
|
It was reported that during a recanalization procedure, the helix was allegedly snapped and come apart from the catheter.There was no reported patient injury.
|
|
Event Description
|
It was reported that during a recanalization procedure in the right superficial femoral artery via right common femoral, the helix was allegedly snapped and come apart from the catheter.There was no reported patient injury.
|
|
Manufacturer Narrative
|
H10: manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the sample was not returned to the manufacturer for evaluation.The investigation is inconclusive for reported helix break issue.Clear root cause could not be identified but a helix break represents a known inherent risk.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: the medical device manufacturer (d3) and manufacturing location (g1) for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.H10: b5, d4 (expiry date: 09/2023), g3, h6 (method) h11: g1, h6 (result, conclusion) h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
|
Event Description
|
It was reported that during a recanalization procedure in the right superficial femoral artery via right common femoral, the helix was allegedly snapped and come apart from the catheter.There was no reported patient injury.
|
|
Manufacturer Narrative
|
H10: manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the sample was returned to the manufacturer for evaluation.A physical investigation was performed for the catheter.The helix was broken at 115 cm from the tip of the catheter.The investigation is confirmed for reported helix break issue.A clear root cause could not be identified but a helix break represents a known inherent risk.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: the medical device manufacturer (d3) and manufacturing location (g1) for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.H10: d4 (expiry date: 09/2023).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|
|