Model Number 106015
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/30/2022
Event Type  malfunction  
Manufacturer Narrative
Manufacturer's investigation conclusion: the relevant sections of the device history records were reviewed and showed no deviation from manufacturing or quality assurance specifications.The heartmate ii left ventricular assist system instructions for use discusses damage due to wear and fatigue of the driveline and includes driveline care instructions.The heartmate ii left ventricular assist system patient handbook also contains information on caring for the driveline.The reported separation of the driveline¿s external silastic sleeve from the controller metal connector was confirmed via an abbott representative.A clamshell repair was performed on (b)(6) 2022 to remedy the situation, per the account.No further information was provided.The manufacturer is closing the file on this event.
 
Event Description
It was reported that the patient's bend relief separated from the metal connector.A universal percutaneous lead bend relief clam shell was installed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTMATE II LVAS IMPLANT KIT (WITH SEALED GRAFTS)
Type of Device
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer (Section G)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer Contact
bob fryc
6035 stoneridge drive
pleasanton, CA 94588
7818528204
MDR Report Key14186275
MDR Text Key289870397
Report Number2916596-2022-10133
Device Sequence Number1
Product Code DSQ
UDI-Device Identifier00813024011224
UDI-Public00813024011224
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P060040
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/22/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2018
Device Model Number106015
Device Catalogue Number106015
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/30/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/18/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age77 YR
Patient Weight90 KG
-
-