| Model Number |
4515-70-000 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Unspecified Tissue Injury (4559)
|
| Date of Event |
04/04/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Surgeon started velys ras attune ps primary knee after completion of all initialization, calibration and draping as per the ifu.Natural joint assessment and planning was then completed and the surgeon chose to perform a 2 degree varus cut on the tibia.The knee was positioned with 'z' retractors placed medial and lateral to protect the collateral ligaments.After finishing the cut with an osteotome the sensor tensor was then used to check balance with the accubalance graph.It was then noted an unexpected increase in medial laxity relative to the lateral compartment.After further assessment it was identified that the mcl was damaged and need repair with fibrewire.This was completed and the femoral cuts were completed with navigation.The rest of the case proceeded without further delay or incident.Doe: (b)(6) 2022.
|
| |
|
Event or Problem Description
|
|
There was a delay of 10 minutes.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : an analysis of the product could not be performed since a physical sample was not received for evaluation.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product or product id numbers are received at a later date, the investigation will be updated as applicable.
|
| |
|
Search Alerts/Recalls
|