| Model Number |
21-4455-24 |
| Medical Device Problem Code |
Device Damaged Prior to Use (2284)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/02/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the catheter scuffed at tip and not conically shaped as usual.Normally, the end to be implanted is conically ground/shaped.The catheter was not implanted and a new port from a different lot was applied.No injury.
|
| |
|
Additional Manufacturer Narrative
|
|
This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).No problems or issues were identified during this device history record review.A product sample was received for evaluation.Visual and functional testing were performed.The returned sample was received inside of a plastic bag without its original packaging.Two catheters were included in the sample returned.It was observed that the tip was not totally chamfered, for one of the catheters and the rest of the components and assemblies no damages nor workmanship defects were detected.The root cause of the reported issue was found to be the most probable root cause of failure mode reported was defined per operation was not correctly performed since catheter tip is partially chamfered, catheter was not rotated during tip forming.Actions were taken to mitigate the reported issue: an awareness notification was issued to production personnel in order to explain the importance to adherence or following in the procedure.
|
| |
|