MENTOR TEXAS MENTOR MEMORYSHAPE¿ TM+ 445CC; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Model Number |
3341309 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Codes |
Deformity/ Disfigurement (2360); Wrinkling (2613)
|
| Date of Event |
03/23/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: wrinkling, anistomasia, rupture.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a (b)(6) caucasian female patient who underwent a breast augmentation revision with mentor memoryshape, 445cc silicone breast implant has experienced right-sided implant rupture, bilateral wrinkling, and bilateral ptosis postoperatively.The report indicated the following: gel rupture, shape, size, rippling and pre-pectoral position.The rupture was discovered during surgery.As a result, patient underwent bilateral mastopexy and, replacements as follow: l/ cat#: smpx270, (b)(4) r/ cat#: smpx270, (b)(4) on (b)(6) 2022.This report relates to the right prosthesis.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information received on april 22, 2022 indicated that the patient has two other event with capsular contracture which will be reported accordingly.The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluations' image completed on may 18, 2022: upon visual evaluation of the image provided in the complaint, the implant was observed ruptured.Some breast ptosis is a normal component of the mature breast.Ptosis becomes undesirable when the breast parenchyma predominates below the areola, droops considerably below the inframammary fold and the nipple points downward.Augmentation alone, without consideration of the ptosis can produce a less than desirable cosmetic result known as a "rock in sock" deformity or an increased ptotic appearance.Ptosis is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of mentor¿s quality process all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.,according to the information received, the patient experienced a rupture in the mentor memoryshape¿ tm+ 445cc breast implant.Additionally, developed breast ptosis, rippling, and pre-pectoral position.Device evaluation completed on may 24 , 2022: the product was returned to mentor for evaluation.Mentor conducted a visual inspection and microscopic examination of the returned device.Visual analysis of the returned sample revealed that the mentor memoryshape¿ tm+ 445cc breast implant was found to be ruptured and received in two parts.A microscopic examination was performed, and the cause of the tear could not be identified.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: rupture can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.The following things may cause the implant to rupture: damage by surgical instruments; stressing the implant during implantation and weakening it; folding or wrinkling of the implant shell; excessive force to the chest (e.G.During closed capsulotomy); trauma; compression during mammographic imaging; and severe capsular contracture.Breast implants may also simply wear out over time.Some breast ptosis is a normal component of the mature breast.Ptosis becomes undesirable when the breast parenchyma predominates below the areola, droops considerably below the inframammary fold, and the nipple points downward.Augmentation alone, without consideration of the ptosis, can produce a less than desirable cosmetic result known as a "rock in a sock" deformity or an increased ptotic appearance.Ptosis is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Manufacturer¿s reference number: (b)(4).
|
| |
|