• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION STERLING; CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION STERLING; CATHETER, PERCUTANEOUS Back to Search Results
Model Number 24700
Medical Device Problem Codes Deflation Problem (1149); Inflation Problem (1310)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/31/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that deflation failure occurred.The target lesion was located in left cephalic vein.A 5.0mmx40mmx135cm sterling balloon catheter was advanced for dilation.The balloon was held to 10 atmospheres, however, it would not inflate.When it was removed from the body, it was noted to be fully deflated and when attempted to inflate outside the body, it did finally inflate, but it would not deflate at all.The procedure was completed with a 5x40 non-boston scientific device.There were no patient complications nor injuries reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of a sterling balloon catheter.The outer shaft, inner shaft, balloon, and tip were visually and microscopically examined.The catheter batch number is a match to the complaint batch number.Visual examination showed a kink 95 cm from the hub.Microscopic examination revealed a kink at the marker band under the balloon.The device was inflated but it took about a minute to fully inflate.When deflating the device it took approximate 1 minute.The account may not have been in a position to wait for the device to deflate.Device analysis determined the condition of the returned device was not consistent with the reported information of failure to inflate and failure to deflate.
 
Event or Problem Description
It was reported that deflation failure occurred.The target lesion was located in left cephalic vein.A 5.0mmx40mmx135cm sterling balloon catheter was advanced for dilation.The balloon was held to 10 atmospheres, however, it would not inflate.When it was removed from the body, it was noted to be fully deflated and when attempted to inflate outside the body, it did finally inflate, but it would not deflate at all.The procedure was completed with a 5x40 non-boston scientific device.There were no patient complications nor injuries reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERLING
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key14187128
Report Number2134265-2022-04072
Device Sequence Number12245406
Product Code DQY
UDI-Device Identifier08714729122906
UDI-Public08714729122906
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K053116
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/25/2024
Device Model Number24700
Device Catalogue Number24700
Device Lot Number0026870913
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/07/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/31/2022
Supplement Date Received by Manufacturer05/05/2022
Initial Report FDA Received Date04/22/2022
Supplement Report FDA Received Date05/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/25/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age54 YR
Patient SexMale
Patient RaceWhite
-
-