MEDTRONIC PUERTO RICO OPERATIONS CO. 630G INSULIN PUMP MMT-1715KL 630G; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-1715KL
Device Problem Human-Device Interface Problem (2949)
Patient Problem Hyperglycemia (1905)
Event Date 08/14/2021
Event Type  Injury  
Manufacturer Narrative
This report is part of a retrospective review and remediation efforts in response to a warning letter.(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event Description
It was identified that the customer's basal programming was unknown.The customer reported that they had high blood glucose since they started the replacement insulin pump.The customer's blood glucose at the time of the event was 600 mg/dl.The customer went to the hospital and her blood glucose was reduced to 280 mg/dl.Troubleshooting was performed and the customer indicated that currently she had no symptoms of high blood glucose.The customer had been using the insulin pump within 48 hours of the high blood glucose event.The insulin pump will not be returned for analysis.
 
Event Description
It was identified that the customer's basal programming was unknown.
 
Manufacturer Narrative
This report is part of a retrospective review and remediation efforts in response to a warning letter.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Additional information has been received which was not included with the initial report.
 
Manufacturer Narrative
The test p-cap locks properly in place in the reservoir compartment noted.Pump received with pillowing keypad overlay during the visual inspection.Thus and carelink software was utilized and downloaded trace/history files properly.Pump passed the self test, sleep current measurement, active current measurement, rewind test, prime/seating test, basic occlusion test, occlusion test, force sensor test, displacement test and dat at 0.08710 inches.Pump programmed with a basal rate of 2.0u/hr and monitored.All basal rates registered properly in the pump history summary noted.Pump was cut open to perform visual inspection and found no moisture or component damage on the electronics, force sensor and motor assembly noted.The force sensor offset measured (23.6 mv).The motor was tested outside of the device on the ngp stb3 and passed.In summary, pump passed all required testing.Unable to verify customer alleged for high bg.The force sensor is within specification and the motor functioning properly.Customer alleged not confirmed for basal rate issue.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
630G INSULIN PUMP MMT-1715KL 630G
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
tricha miles
ceiba norte ind. park #50 road
juncos 00777--386
7635140379
MDR Report Key14187476
MDR Text Key289878832
Report Number2032227-2022-182705
Device Sequence Number1
Product Code OZO
UDI-Device Identifier000000763000367053
UDI-Public(01)000000763000367053(17)240714
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 07/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date07/14/2024
Device Model NumberMMT-1715KL
Device Catalogue NumberMMT-1715KL
Device Lot NumberHG5H5QB
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/21/2021
Initial Date FDA Received04/22/2022
Supplement Dates Manufacturer Received04/27/2022
07/21/2022
Supplement Dates FDA Received04/27/2022
07/26/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/15/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberZ-0855-2022
Patient Sequence Number1
Patient Age76 YR
Patient SexFemale
-
-