Model Number DAT315
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Visual Impairment (2138)
Date of Event 11/01/2021
Type of Reportable Event Serious Injury
Event or Problem Description
A purchasing manager reported that following an intraocular lens (iol) implant procedure, the patient had experienced poor visual acuity.The iol was exchanged for another lens approximately 5 months following the initial implant procedure.Additional information was requested.
 
Additional Manufacturer Narrative
The explanted lens was returned in a lens case inside an opened non company's lens carton.Viscoelastic and blood was observed on the lens.One haptic was bent toward the optic edge.The optic was cut into two portions, typical in appearance to damage created to facilitate removal from the eye.The posterior optic surface has a line of small cracks.Product history records were reviewed and documentation indicated the product met release criteria.Associated product information was not provided.It is unknown if qualified products were used.An iol exchange for a difference equal to or greater than two diopters, indicates the event is most likely related to a calculation error.Based on the information provided, the event does not appear to be related to the product.Information was provided that the 18.5 diopter lens was exchanged for a non-company lens model that was three diopters higher.This information may suggest that the replacement lens was a better selection for the patient's vision needs or preferences.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ VIVITY TORIC EXTENDED VISION UV ABSORBING IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14187679
Report Number1119421-2022-00826
Device Sequence Number4481890
Product Code POE
UDI-Device Identifier00380652427067
UDI-Public00380652427067
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/29/2025
Device Model NumberDAT315
Device Lot Number15085728
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/12/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/31/2022
Initial Report FDA Received Date04/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age86 YR
Patient SexMale
-
-