• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TORNIER S.A.S. AEQUALIS REVERSED MULTIDIR. SCREW D4.5 L 38 STERILE; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORNIER S.A.S. AEQUALIS REVERSED MULTIDIR. SCREW D4.5 L 38 STERILE; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Model Number DWD138
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Codes Bacterial Infection (1735); Metal Related Pathology (4530); Implant Pain (4561)
Date of Event 11/23/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient attended a revision surgery for a reverse ii implant, implanted in 2014.The patient reported with pain, loosening and infection.It's possible that the infection was caused by metalosis due to implant loosening.
 
Additional Manufacturer Narrative
Device will not be returned.If additional information becomes available, it will be provided on a supplemental report.
 
Event or Problem Description
It was reported that the patient attended a revision surgery for a reverse ii implant, implanted in 2014.The patient reported with pain, loosening and infection.It's possible that the infection was caused by metalosis due to implant loosening.
 
Additional Manufacturer Narrative
The reported event could be confirmed, since the micro culture and pathology results confirm the infection of the patient by pseudomonas aeruginosa.This conclusion was verified by a medical expert upon review of the patient's case documentation.Medical expertise : "indeed, from the attached documentation all the evidence points in one direction, that of an infected implant, with positive cultures showing the same bacteria from swabs taken during the first re-operation (surgical debridement and deep swabs) on (b)(6) 2021, and the tissue samples taken during the revision surgery on (b)(6) 2021.The infection was caused by pseudomonas aeruginosa, a gram-stain negative bacteria." moreover, the device inspection did not reveal any factors that could have led the reported problem.The root cause could not be determined.However, based on the clinical opinion on the potential risk of infection caused by sterile packed stryker implants and instruments, ¿any kind of surgery bears the risk of infection.In trauma and orthopedic surgery with external or internal fixation implants or endoprostheses the infection risk varies between 0.5 % and 5 %, exceptionally up to 10 % and beyond, mainly depending on the respective surgical procedure and the typical patient population." a review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If more information is provided, the case will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AEQUALIS REVERSED MULTIDIR. SCREW D4.5 L 38 STERILE
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
TORNIER S.A.S.
161 rue lavoisier
montbonnot saint-martin 38330
FR  38330
Manufacturer (Section G)
TORNIER S.A.S.
161 rue lavoisier
montbonnot saint-martin 38330
FR   38330
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key14188150
Report Number3000931034-2022-00207
Device Sequence Number4378821
Product Code KWS
UDI-Device Identifier03700386917110
UDI-Public03700386917110
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K081059
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/30/2019
Device Model NumberDWD138
Device Catalogue NumberDWD138
Device Lot Number2062AO
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/28/2022
Supplement Date Received by Manufacturer05/11/2022
Initial Report FDA Received Date04/22/2022
Supplement Report FDA Received Date06/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/23/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age85 YR
Patient SexMale
-
-