COVIDIEN MFG SOLUTIONS S.A. MAHURKAR; CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS
|
Back to Search Results |
|
| Model Number |
8831661001 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/02/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, during treatment, the venous end of the device was found to be prolonged and damaged (hole).There was a leak at the venous exit site.The treatment was completed after the device was replaced, there was no blood loss and blood transfusion was not required.There was no intervention/treatment required for the leakage, nothing unusual was observed on the device prior to use and flushing was performed with normal result.There was no other product utilized with the device, iodophor disinfection was used to treat the insertion site prior to product placement and there was no medical intervention done to the patient.The clamps were move to a new position after each treatment, there was no cleaning agent used to clean the adapters and the cleaning agent was allowed to dry thoroughly prior to applying ointment to the area.The catheter was not repaired, there was no cleaning agent used on the device, tego was not utilized and there was no luer adapter issue.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: a5a, a5b, b5, g3 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during treatment, the venous end of the device was found to be prolonged and damaged (hole).There was a leak at the venous exit site.There was nothing unusual was observed on the device prior to use and there was no other product utilized with the device.Flushing was performed with normal result and iodophor disinfection was used to treat the insertion site prior to product placement.The clamps were moved to a new position after each treatment, there was no cleaning agent used to clean the adapters and the cleaning agent was allowed to dry thoroughly prior to applying ointment to the area.The catheter was not repaired, there was no cleaning agent used on the device, tego was not utilized and there was no luer adapter issue.The treatment was completed after the device was replaced.There was no blood loss and blood transfusion was not required.There was no intervention/treatment required for the leakage and there was no medical intervention done to the patient.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g1(manufacturer name, mfr contact first name, last name, street 2, mfr city, region, postal code, email and phone number), g3, h3, h6 correction: pt.Identifier, age at event, sex, and weight in lbs were removed, e1(facility name) h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection noted there was a cut in the extension tube where it met the venous luer adapter.The placement of the cut suggests it was created by clamping the tubing too close to the luer adapter.Clamping the extension tube close to the luer adapter can cause excess stress on the extension tubing which can lead to breaks or leak in the tubing where it connects to the luer adapter.It was reported that there was a leak on the extension tube at or near the luer adapter.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: clamping repeatedly on the same spot could weaken the tubing and clamping near the adapter and hub should be avoided.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|