Model Number N/A
Medical Device Problem Code Difficult or Delayed Separation (4044)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/30/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of refx4452 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported "introducer didn't break completely and had to do guidewire exchange.Possible loss of vessel and patient with very limited venous access.".
 
Event or Problem Description
It was reported "introducer didn't break completely and had to do guidewire exchange.Possible loss of vessel and patient with very limited venous access.".
 
Additional Manufacturer Narrative
The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample analysis, applicable fmea documents, applicable manufacturing records, and labeling.Based on a review of this information, the following was concluded: the complaint of an improper peel is confirmed but the exact cause remains unknown.Six photo samples of a 4.5 fr micoez microintroducer were provided for evaluation.The photos all showed the proximal section of the microintroducer containing the orange tabs.The tabs and grey sheath are partially peeled with blood residue present on the device surface.The catheter was present within the orange tabs and grey sheath.The catheter appeared to be retained by a full ring on one half of the split orange tab.Since the tab peel appears to be unevenly split, the complaint is confirmed but the exact factors remain unknown.Based on the information provided possible contributing factors include material properties and uneven force applied during splitting of the tabs.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOT BARD POLY MIDLINE
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
johanna de oliveira
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key14188456
Report Number3006260740-2022-01408
Device Sequence Number14591619
Product Code LJS
UDI-Device Identifier00801741121609
UDI-Public(01)00801741121609
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K001901
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberCK000495A
Device Lot NumberREFX4452
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/30/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/30/2022
Supplement Date Received by Manufacturer05/26/2022
Initial Report FDA Received Date04/22/2022
Supplement Report FDA Received Date06/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-