• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWER-TRIALYSIS SHORT-TERM STRAIGHT DIALYSIS CATHETER; CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWER-TRIALYSIS SHORT-TERM STRAIGHT DIALYSIS CATHETER; CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED Back to Search Results
Model Number N/A
Medical Device Problem Code Material Frayed (1262)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/21/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of refy2001 showed one other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported via medwatch "provider and the registered nurse attempted a right femoral placement of the power trialysis short term straight dialysis catheter.They aborted the attempt due to equipment failure.While trying to put in a power-trialysis catheter, the guide wire came apart, outside of the patient's body (not inside).They had to stop the procedure because they could not thread the catheter over to the guide wire.At that point there was not harm to the patient other than the aborted attempt, and some bleeding during the procedure.The wire was kept.Time, date and patient label were placed on the clear plastic bag with the guidewire and introducer catheter, that caused the failure." add info rcvd 04/11/2022: placement date: (b)(6) 2022, explant date: (b)(6) 2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWER-TRIALYSIS SHORT-TERM STRAIGHT DIALYSIS CATHETER
Common Device Name
CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
kayla olsen
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key14188590
Report Number3006260740-2022-01413
Device Sequence Number19916610
Product Code NIE
UDI-Device Identifier00801741086960
UDI-Public(01)00801741086960
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K133456
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number5605200
Device Lot NumberREFY2001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/04/2022
Initial Report FDA Received Date04/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age56 YR
Patient SexFemale
-
-