| Model Number |
N/A |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Material Deformation (2976)
|
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/30/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of refy3552 showed two other similar product complaint(s) from this lot number.
|
| |
|
Event or Problem Description
|
|
It was reported "complaints from complex care of cracking catheter hub when connected to dead end cap result in patient needing catheter replacement.".
|
| |
|
Additional Manufacturer Narrative
|
|
The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample analysis, manufacture quality and production records, and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a leaking catheter was confirmed, and the cause was manufacturing-related.The product returned for evaluation was one powerpicc 5 fr s/l ir catheter.Light usage residues were observed throughout the sample and a needleless injection cap was attached to the luer adapter.Microscopic inspection of the luer adapter revealed a protrusion in the luer material within the orifice.That material had the appearance of a mold defect which could result in inadequate sealing of the luer hub with a syringe.Bd is working closely with manufacturing to prevent recurrence of the reported event.
|
| |
|
Event or Problem Description
|
|
It was reported "complaints from complex care of cracking catheter hub when connected to dead end cap result in patient needing catheter replacement.".
|
| |
|