C.R. BARD, INC. (BASD) -3006260740 POWERPICC 5F SINGLE-LUMEN POLYURETHANE CATHETER RADIOLOGY 70 CM GW; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Medical Device Problem Codes Fluid/Blood Leak (1250); Material Deformation (2976)
Health Effect - Clinical Codes Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/30/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of refy3552 showed two other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported "complaints from complex care of cracking catheter hub when connected to dead end cap result in patient needing catheter replacement.".
 
Additional Manufacturer Narrative
The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample analysis, manufacture quality and production records, and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a leaking catheter was confirmed, and the cause was manufacturing-related.The product returned for evaluation was one powerpicc 5 fr s/l ir catheter.Light usage residues were observed throughout the sample and a needleless injection cap was attached to the luer adapter.Microscopic inspection of the luer adapter revealed a protrusion in the luer material within the orifice.That material had the appearance of a mold defect which could result in inadequate sealing of the luer hub with a syringe.Bd is working closely with manufacturing to prevent recurrence of the reported event.
 
Event or Problem Description
It was reported "complaints from complex care of cracking catheter hub when connected to dead end cap result in patient needing catheter replacement.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC 5F SINGLE-LUMEN POLYURETHANE CATHETER RADIOLOGY 70 CM GW
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
johanna de oliveira
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key14189477
Report Number3006260740-2022-01418
Device Sequence Number4406641
Product Code LJS
UDI-Device Identifier00801741027697
UDI-Public(01)00801741027697
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K033389
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number3175335
Device Lot NumberREFY3552
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/30/2022
Supplement Date Received by Manufacturer06/01/2022
Initial Report FDA Received Date04/22/2022
Supplement Report FDA Received Date06/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;