• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA GMK-REVISION TIBIAL INSERT PS FIXED SIZE 2/12MM; KNEE FIXED TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA GMK-REVISION TIBIAL INSERT PS FIXED SIZE 2/12MM; KNEE FIXED TIBIAL INSERT Back to Search Results
Model Number 02.07.0212PSF
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Pain (1994)
Date of Event 03/23/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on (b)(6) 2022, lot 168560: 30 items manufactured and released on (b)(6) 2017.Expiration date: (b)(6) 2022.No anomalies found related to the problem.To date, all items of the same lot have been sold with no similar reported event.Clinical evaluation performed by medical affairs director at 2 years after a complex revision tka surgery, a traumatic event occurs that reportedly determined the fracture of the insert fixation screw and partial dislocation of the insert.The report does not specify if a new screw could be implanted, and no post-revision xrays are supplied.It is conceivable that no screw could be used for the second revision.It must be noted that the original insert was of low thickness (12mm), there is no information as to the thickness of the new insert used.The lower the insert, the more stable it can be expected to be, even in absence of the fixation screw.Given the complex situation of the involved joint, removal/substitution of the tibial tray should be avoided, and most likely the condition of minimum damage to the patient is to leave the existing tibial tray and the insert clipped with no additional fixation screw.However, the behaviour of the prosthesis in these conditions cannot be predicted.
 
Event or Problem Description
The patient came in reporting pain after falling.About 1 year and 9 months after the primary surgery, the surgeon observed that the screw in the patient's poly had broken and that the poly had partially dislocated.The surgeon removed the broken screw and revised the poly.The surgery was completed successfully.3.5nm torque limiter used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GMK-REVISION TIBIAL INSERT PS FIXED SIZE 2/12MM
Common Device Name
KNEE FIXED TIBIAL INSERT
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ   6874
Manufacturer Contact
stefano baj
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key14189887
Report Number3005180920-2022-00279
Device Sequence Number4481410
Product Code JWH
UDI-Device Identifier07630030816581
UDI-Public07630030816581
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K090988
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/23/2022
Device Model Number02.07.0212PSF
Device Catalogue Number02.07.0212PSF
Device Lot Number168560
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/23/2022
Initial Report FDA Received Date04/22/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/14/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Disability; Required Intervention;
Patient SexFemale
-
-