| Model Number |
RLT351418 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Aneurysm (1708)
|
| Date of Event |
04/14/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Patient medications: tamsulosin, midodrine, and aspirin.
|
| |
|
Event or Problem Description
|
|
The field sales associate (fsa) reported the following: on (b)(6) 2020, the patient was implanted with a gore® excluder® aaa endoprostheses to treat an abdominal aortic aneurysm.The patient tolerated the procedure.It was reported that images (modality and date is unknown), revealed a proximal type i endoleak.It is unknown if there is aneurysm sac enlargement.Reportedly, during the initial procedure the gore® excluder® trunk ipsilateral leg endoprosthesis was intentionally deployed lower distally to the renal arteries due to a reverse tapered aortic neck.There was a reintervention on (b)(6) 2022.The physician implanted two gore® excluder® aaa endoprostheses aortic extender in the proximal seal zone from the original procedure.The endoleak was resolved.The patient tolerated the reintervention procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Addition code 22: the gore® excluder® aaa endoprosthesis instructions for use (ifu) states that adverse events that may occur and/or require intervention include but are not limited to endoleak.
|
| |
|