| Model Number |
3500 |
| Medical Device Problem Code |
Device Alarm System (1012)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Information was received indicating that the pump failed the occlusion test due to "system failure: primary audible alarm background test." it is unknown if there was any patient involvement or any adverse effects.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was received indicating there was no patient involvement (updated h6 event problem patient code).
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to smiths/icu medical.There were no signs of external damage.The error history log cited a primary audible alarm bgnd test." the reported problem could not be duplicated.The speaker was replaced as a preventative measure.No causes or potential causes to the customer's reported problem were found during the dhr review.
|
| |
|