• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS IOL; LENS, INTRAOCULAR, TORIC OPTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS IOL; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number DIU225
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/30/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Age, weight and ethnicity: asked but information not provided.If explanted, give date: not applicable, as lens remains implanted.The device was not returned for analysis as it remains implanted.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the simplicity intraocular lens (iol) has a scuff after being inserted into the patient's capsular bag.The lens remains implanted.No further information was provided.
 
Additional Manufacturer Narrative
Device evaluation: a picture provided in the complaint file and attached to the current document was evaluated, customer provided picture shows a pseudophakic eye, implanted with an intraocular lens claimed to be a tecnis eyhance toric ii one-piece iol with simplicity delivery system model diu225.There are observed 4 small aligned marks located on the optical surface near to the lens periphery at 2:00.The root cause and the definitive clinical impact for the observed phenomena, cannot be determined from a picture assessment.The complaint issue "dc-cosmetic issues" was identified during product evaluation; however, based on the complaint investigation results the complaint issue could not be confirmed to be related to the manufacturing or design process.Therefore, there is no indication of a product deficiency or product malfunction.Conclusion: as a result of the investigation there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
1700 e st andrew place
santa ana, CA 92705
7142478552
MDR Report Key14190219
Report Number3012236936-2022-01078
Device Sequence Number11845369
Product Code MJP
UDI-Device Identifier05050474747111
UDI-Public(01)05050474747111(17)240725
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/25/2024
Device Model NumberDIU225
Device Catalogue NumberDIU225U190
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/30/2022
Supplement Date Received by Manufacturer07/16/2022
Initial Report FDA Received Date04/22/2022
Supplement Report FDA Received Date07/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/25/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-