| Model Number |
24700 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial reporter title: (b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that a balloon issue occurred.A 3.0mm x 40mm x 90cm sterling over-the-wire balloon catheter was selected for a percutaneous transluminal angioplasty procedure in the superficial femoral artery.During the procedure, the balloon was separated from the shaft.The balloon was removed.The procedure was completed with another of the same device.There were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Initial reporter title: buyer device eval by manufacturer: the returned device consisted of a sterling balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual and microscopic examination revealed no damages.Inspection of the remainder of the device showed no other damage or irregularities.Product analysis could not confirm the reported separation.
|
| |
|
Event or Problem Description
|
|
It was reported that a balloon issue occurred.A 3.0mm x 40mm x 90cm sterling over-the-wire balloon catheter was selected for a percutaneous transluminal angioplasty procedure in the superficial femoral artery.During the procedure, the balloon was separated from the shaft.The balloon was removed.The procedure was completed with another of the same device.There were no patient complications reported.
|
| |
|
Search Alerts/Recalls
|