MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION 2CM PERIPHERAL CUTTING BALLOON; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION 2CM PERIPHERAL CUTTING BALLOON; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA Back to Search Results
Model Number 24628
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: used (b)(6) 2022 as the event date was not reported.
 
Event or Problem Description
It was reported that the blade was lifted.A 5.00mm/2.0cm/50cm peripheral cutting balloon was selected for use.During the procedure, it was noted that the cutting wire broke free.No patient complications were reported.
 
Event or Problem Description
It was reported that the blade was lifted.A 5.00mm/2.0cm/50cm peripheral cutting balloon was selected for use.During the procedure, it was noted that the cutting wire broke free.No patient complications were reported.
 
Additional Manufacturer Narrative
B3 - date of event: used 03/01/2022 as the event date was not reported.Device evaluated by mfr: the device was returned for evaluation.A visual examination of the returned device identified that the balloon had been inflated and was not refolded.It is not known when the balloon was subjected to positive pressure.The balloon of the device was visually examined, and no issues were noted that may have potentially contributed to the complaint incident.A visual examination of the returned device identified that a 10mm section of blade was noted to have lifted.The blade pad and all other blades were present and fully bonded to the balloon surface.No issues were noted with the tip section of the device.A visual examination found no issue with the markerbands.A visual and tactile examination identified a shaft kink 35mm distal from the distal end of the strain relief.No other issues were identified during the product analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2CM PERIPHERAL CUTTING BALLOON
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key14190632
Report Number2134265-2022-04604
Device Sequence Number4378914
Product Code PNO
UDI-Device Identifier08714729813705
UDI-Public08714729813705
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K151253
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/28/2022
Device Model Number24628
Device Catalogue Number24628
Device Lot Number0026260258
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/21/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/31/2022
Supplement Date Received by Manufacturer04/29/2022
Initial Report FDA Received Date04/22/2022
Supplement Report FDA Received Date05/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/28/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-