ABBOTT MEDICAL AMPLATZER VASCULAR PLUG II; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number 9-AVP2-006
Medical Device Problem Codes Migration or Expulsion of Device (1395); Off-Label Use (1494)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/24/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on (b)(6) 2022, a 6mm amplatzer vascular plug ii was chosen for implant using a non-abbott delivery system.The device was sized using angiogram measurements that were taken.During procedure, the plug was implanted into the patent ductus arteriosus (pda).Once the device was released, the proximal portion of the device shifted to the main pulmonary artery causing an obstruction of flow.The device was attempted to be retrieved, but it was not able to be retrieved in the catheterization laboratory.The patient was referred to surgery for removal of the device and closure of the pda.The patient remained hemodynamically stable throughout the event.The patient status was reported to be stable.The cause of the migration was unknown.No additional information was provided.
 
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
An event of an amplatzer vascular plug ii being used off-label to close a patent ductus arteriosus and the device migrating to block the pulmonary artery was reported.A returned device assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The use of the device in a position not recommended in the instructions for use could have contributed to the reported event.There is no indication of a product quality issue with regards to manufacture, design, or labeling.Please note, per the instructions for use, "the safety and effectiveness of this device for cardiac uses (for example, patent ductus arteriosus or paravalvular leak closures) and neurological uses have not been established.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER VASCULAR PLUG II
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL REG# 2135147
5050 nathan ln n
plymouth MN 55442
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key14190867
Report Number2135147-2022-00191
Device Sequence Number19935060
Product Code KRD
UDI-Device Identifier00811806010366
UDI-Public00811806010366
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
K071699
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9-AVP2-006
Device Catalogue Number9-AVP2-006
Device Lot Number8321236
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/24/2022
Supplement Date Received by Manufacturer06/20/2022
Initial Report FDA Received Date04/22/2022
Supplement Report FDA Received Date07/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/02/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NON-ABBOTT DELIVERY SYSTEM, UNKNOWN MANUFACTURER.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexFemale
-
-