ABBOTT MEDICAL AMPLATZER SEPTAL OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER Back to Search Results
Model Number 9-ASD-030
Medical Device Problem Code Migration or Expulsion of Device (1395)
Health Effect - Clinical Codes Air Embolism (1697); Arrhythmia (1721); Bradycardia (1751); Stroke/CVA (1770); Embolism/Embolus (4438)
Date of Event 03/25/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a 28mm amplatzer septal occluder was selected for an implanted on (b)(6) 2022 for an atrial septal defect closure procedure.Using a 24mm unknown sizing balloon, the defect was sized at 27mm.The 28mm occluder was prepped and delivered to the defect.Intracardiac echocardiography (ice) showed some insufficient apposition to the aortic root (ao) rim, and when the stability test was performed, the device prolapsed thru the defect.The device was recaptured and removed from the patient prior to release from the delivery cable.A 30mm amplatzer septal occluder was prepped and delivered to the defect.Ice imaging showed better apposition to the rims, with some color doppler flow noted thru the device, but not around either disc.It was noted that three separate stability checks were performed with each one showing the device to be stable on the septum.The device was released from the delivery cable and subsequent ice imaging showed the device to be seated well on the septum.The patient remained stable at the point.Two preclose devices were prepped for deployment, as the first preclose was being deployed, the patient began to experience some ectopy, when the patient was placed under fluoroscopy, the device was noted to have embolized complete into the patient right pulmonary artery (pa).The right pa was not completely occluded.The decision was made to prepare for device retrieval using a large 12 fr sheath and a gooseneck snare.The patient continued to remain stable.After numerous unsuccessful attempts to snare the device, an ensnare was then introduced to try and snare the device after removing the gooseneck snare.Another sheath was placed in the left femoral artery and a pig tail catheter was introduced into the pa to try and steer the device while attempting to snare it.These attempts were also unsuccessful and towards the end of this attempt, a series of air bubbles were noted on the fluoroscopy moving thru the pa.The patient became unstable with stroke/seizure like symptoms, unable to focus or respond along with decreased heart rate and blood pressure.High flow oxygen was given along with atropine and fluids.The patient rapidly regained movement and eye focus and remained somewhat stable.The stroke team was present at this time and the decision was made to take the patient to computerized tomography (ct).The patient remained in the unit with high flow oxygen and heparin.Once stable, the patient was taken to emergency surgery to remove the 30mm amplatzer septal occluder and patch the defect.The event did contribute to a clinically significant delay in procedure.At the time of report, the patient was noted to have been recovering.
 
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
An event of device embolization, ectopy, and stroke/seizure like symptoms was reported.A returned device assessment could not be performed as the device remains in the patient and was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Common Device Name
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL REG# 2135147
5050 nathan ln n
plymouth MN 55442
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key14191078
Report Number2135147-2022-00197
Device Sequence Number11873937
Product Code MLV
UDI-Device Identifier05415067019332
UDI-Public05415067019332
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P000039
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9-ASD-030
Device Catalogue Number9-ASD-030
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/25/2022
Supplement Date Received by Manufacturer05/25/2022
Initial Report FDA Received Date04/23/2022
Supplement Report FDA Received Date06/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Required Intervention;
Patient Age64 YR
Patient SexFemale
-
-