Model Number 02.07.0514SCF
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 03/31/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 12 april 2022.Lot 189547: (b)(4) items manufactured and released on 01-mar-2019.Expiration date: 2024-02-18.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.
 
Event or Problem Description
The patient had a primary knee surgery on (b)(6) 2012.On (b)(6) 2020 the surgeon revised all implants following mobilization.Presently, on (b)(6) 2022, the patient came in reporting instability and the cause is unknown.The surgeon revised the poly and the surgery was completed successfully.The surgeon trialed a 17mm insert but it was not possible to get the leg straight.So, the implanted and the explanted liner have the same height.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GMK-REVISION FIXED TIBIAL INSERT SC SIZE 5/14MM
Common Device Name
KNEE TIBIAL INSERT
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ   6874
Manufacturer Contact
stefano baj
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key14191711
Report Number3005180920-2022-00311
Device Sequence Number17245294
Product Code JWH
UDI-Device Identifier07630030818455
UDI-Public07630030818455
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K103170
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/18/2024
Device Model Number02.07.0514SCF
Device Catalogue Number02.07.0514SCF
Device Lot Number189547
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/31/2022
Initial Report FDA Received Date04/25/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/01/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-