| Model Number |
413419 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Cellulitis (1768); Erythema (1840); Hemorrhage/Blood Loss/Bleeding (1888); Itching Sensation (1943); Tissue Breakdown (2681); Skin Inflammation/ Irritation (4545)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Based on the available information, this event is deemed to be a serious injury.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.(b)(4).
|
| |
|
Event or Problem Description
|
|
The consumer reported to specialist on 31st march 2022 that he experienced skin irritation around tape area after using wafer.Outbound call was made, however, no other information was received.On (b)(6) 2022, consumer called back and reported that he started to have redness on the skin with skin breakdown and bleeding under the tape collar approximately two weeks ago.The redness was only in one area under the tape collar on the side under the stoma.He also had sensations of itching.He went to see his doctor of medicine (md) who felt that he had developed cellulitis under the tape collar and prescribed oral antibiotics.He did not know about antibiotic prescribed by doctor.He also prescribed neosporin to be applied to the area with each wafer change.He completed the course of antibiotics several days ago.The consumer had been using this wafer for at least six months without any difficulties with a wear time of seven days.He reported a skin care routine of using baby shampoo in the shower to clean the peristomal skin.He also uses adhesive remover wipes and skin protectant wipes.The customer was advised to notify his md about the skin condition that had not improved and he continued to have areas of bleeding without blister formation.He denied any evidence of leakage of urine under the wafer.No photo is available at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
Batch record review: lot 1f04731 was manufactured 07/nov/2021, in the ats #2 manufacturing line with a total of (b)(4) mkus.A batch record review was performed on 28/jul/2022, to verify if all the applicable procedures were followed and no issues were found, all the components for assembly were correct per bom and all the tooling information documented was also correct, sap material id (b)(4) and manufacturing order (b)(4).No discrepancies related to the issue reported were found.Complaint data analysis: on 08/26/2022, a search of complaints into trackwise system for lot number 1f04731 was performed.As a result, a total of 01 complaint was found and 01 is related to the same malfunction code of this investigation.Returned sample evaluation: no photographs associated with this case were received and no unused return sample was expected.Conclusion summary of the related event: based on the preliminary investigation performed, no issues related to the customer experience were found in the batch record review.No significant changes have been made in the process or in the product components that could cause the adverse effects reported by the customer.The product was used in their shelf life.No photos or samples were received and per customer description there is not enough information about how the product was used that leads us to a conclusion, therefore, there is not product malfunction confirmed.Should additional information become available, a follow up report will be submitted.Fda registration number: reporting site: 1049092.Manufacturing site: 9618003.
|
| |
|
Event or Problem Description
|
|
To date no additional patient or event details have been received.
|
| |
|
Search Alerts/Recalls
|