Batch review performed on 4 april 2022.Lot 2101520: (b)(4) items manufactured and released on 07-jun-2021.Expiration date: 2026-may-20.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event.Clinical evaluation performed by clinical affairs specialist.Few weeks after a cementless tha, the patient reported a fracture.It was reported that the cerclage was inserted in order to prevent a fracture during femoral preparation.It is conceivable that during surgery, in particular during femoral preparation, the bone may have been weakened and a least-resistance area may have been created.Additionally, during rehabilitation period, a sudden movement on the weakened bone may favour the occurrence of early fracture.This is a known possible adverse event following tha and there is no reason to suspect that it is due to a defect in the implanted device or in the preparing instruments.
|