SMITH & NEPHEW, INC. GNS II CMT TIB SIZE 5 RIGHT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED,
|
Back to Search Results |
|
| Model Number |
71420186 |
| Medical Device Problem Codes |
Device Markings/Labelling Problem (2911); Manufacturing, Packaging or Shipping Problem (2975)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/31/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that, during total knee arthroplasty, a box of genesis ii non-porous tibial baseplate size 5 right contained a size 4 left device instead.The outside of the box displayed part no.71420186 and lot no.20cm03707, but the product inside was a 71420166 from lot no.21fm13870.The incorrect device could be removed before cementing hardened or damage occurred.Surgery was resumed, after a non-significant delay, with the same device.Patient was not harmed as consequence of this problem.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, based on the information provided, the root cause of the reported event was human error; however, the exact point of divergence could not be concluded.The patient impact beyond the reported 0-30 minute surgical extension and modified surgical procedure could not be determined, as it was reported that the procedure resumed using the same device without patient harm.No further medical assessment can be rendered at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for the label specification and manufacturing records revealed that the tibial baseplate size 5 right should be inside the package, matching label and product with part no.71420186 and lot no.20cm03707.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Manufacturing process errors are likely probable causes of the reported event.The contribution of the device to the reported event could not be corroborated.This event was previously identified, a quality hold has been issued as a containment action and a review of the rebox process is being performed to mitigate its reoccurrence.Based on this investigation, the need for corrective action is indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: b5, h6 (medical device problem code, investigation findings and investigation conclusions).Additional information: h7, h11 (roi).H3, h6: the device was not returned for evaluation and the reported event could not be confirmed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.A review of the device labelling, the label specification and manufacturing records revealed that the tibial baseplate size 5 right should be inside the package, matching label and product with part no.71420186 and lot no.20cm03707.The quality team concluded that a manufacturing process error occurred, therefore, the product failed to meet specifications at the time of manufacture.Corrective actions including clarification of rework process procedures, an update to packaging inspection protocols requiring 100% inspection, reinforcement of line clearance in rework instructions, and operator retraining to prevent future mix-ups were implemented.A field action was implemented for distributed product as well.The clinical investigation concluded that the requested clinical information/documentation has not been received as of the date of this medical investigation.Based on the information provided, the root cause of the reported event was human error which resulted in mispacking.The patient impact included the reported 0¿30-minute surgical extension and modified surgical procedure along with use of a s+n back-up component to resume the procedure.No patient harm was reported and no further medical assessment can be rendered at this time.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Event or Problem Description
|
|
It was reported that, during total knee arthroplasty, a box of genesis ii non-porous tibial baseplate size 5 right contained a size 4 left device instead.The outside of the box displayed part no.71420186 and lot no.20cm03707, but the product inside was a 71420166 from lot no.21fm13870.The incorrect device could be removed before cementing hardened or damage occurred.Surgery was resumed, after a non-significant delay, with a s+n back-up device.Patient was not harmed as consequence of this problem.
|
| |
|