Model Number 75004650
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/29/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, during total hip replacement, a polarstem stem inserter broke inside the patient.Surgery was resumed, without any delay, with a smith & nephew back-up device.Patient was not injured as consequence of this problem.
 
Additional Manufacturer Narrative
H3, h6: it was reported that, during total hip replacement, a polarstem stem inserter broke inside the patient.Surgery was resumed, without any delay, with a smith & nephew back-up device.The device, intended for use in treatment, was not returned for investigation.No batch number was communicated, therefore it cannot be determined whether the part met the manufacturing specification at the time of distribution.A complaint history review was conducted.The complaint is in line with the current risk management.According to document "processing (cleaning, disinfection and sterilization) of instruments from smith & nephew orthopaedics ag" (lit.N°03389-en 1363 v3 11/19), all instruments must be inspected and controlled for proper functioning after cleaning/disinfection.Review of past corrective actions was performed.No further escalation is required.Due to insufficient information, the relationship between the reported event and the device cannot be confirmed.The root cause cannot be determined and we cannot speculate about factors which would have contributed to the reported issue.Should the part be returned for investigation, the investigation will be reopened.No further actions are deemed necessary.Smith and nephew will monitor this device for further similar issues.
 
Additional Manufacturer Narrative
It was reported that, during total hip replacement, a polarstem stem inserter broke inside the patient.Surgery was resumed, without any delay, with a smith & nephew back-up device.The device, intended for use in treatment, was not returned for investigation.No batch number was communicated, therefore it cannot be determined whether the part met the manufacturing specification at the time of distribution.Nevertheless, a complaint history review was conducted.The occurrence of the reported failure mode is within its expected risk management level.Furthermore, a risk management review was performed.The complaint is in line with the current risk management.According to document "processing (cleaning, disinfection and sterilization) of instruments from smith & nephew orthopaedics ag", all instruments must be inspected and controlled for proper functioning after cleaning/disinfection.Review of past corrective actions was performed.No further escalation is required.Due to insufficient information, the relationship between the reported event and the device cannot be confirmed.The root cause cannot be determined and we cannot speculate about factors which would have contributed to the reported issue.Should the part be returned for investigation, the investigation will be reopened.No further actions are deemed necessary.Smith and nephew will monitor this device for further similar issues.Internal complaint reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLARSTEM STEM INSERTER
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau TN CH-50 00
SZ  CH-5000
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau TN CH-50 00
SZ   CH-5000
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14192161
Report Number9613369-2022-00225
Device Sequence Number14554599
Product Code LXH
UDI-Device Identifier07611996097573
UDI-Public07611996097573
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number75004650
Device Catalogue Number75004650
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/30/2022
Supplement Date Received by Manufacturer06/01/2022
07/15/2022
Initial Report FDA Received Date04/25/2022
Supplement Report FDA Received Date06/23/2022
08/03/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-