• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CORPORATION, ASHITAKA RADIFOCUS GUIDEWIRE OR GLIDEWIRE OR GLIDEWIRE M; WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CORPORATION, ASHITAKA RADIFOCUS GUIDEWIRE OR GLIDEWIRE OR GLIDEWIRE M; WIRE, GUIDE, CATHETER Back to Search Results
Model Number N/A
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/31/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Implanted date: device was not implanted.Explanted date: device was not explanted.Occupation: ir (interventional radiology) tech.The actual device was not returned for evaluation.The investigation is currently ongoing.A follow-up report will be submitted once the investigation is complete.A review of the manufacturing record and the shipping inspection record of the involved product/lot number combination confirmed there was no problem in them.A search of the complaint file found no other report with the involved product code/lot number combination from other facilities.(b)(4).
 
Event or Problem Description
The user facility reported that the coating stripped off the angled stiff wire involved.Additional information was received on 04 apr 2022: the patient was not affected.Additional information was received on 15 apr 2022: the patient was not injured.Medical or surgical intervention was not required and there was no blood loss.The procedure performed was a lower extremity (le) angiogram.The patient was in stable condition.The procedure outcome was successful.The wire coating stripped off the wire after the quickcross was exchanged.
 
Additional Manufacturer Narrative
This report is being submitted as follow up no.1 to provide the completed investigation results.The actual sample was not returned, therefore no analysis was performed.Based on our past findings, it is known that the outer layer of the radifocus guidewire m may peel off when it is used in combination with a metal needle.Based on the results of the investigation, no anomaly was found in the manufacturing related records.In this case, it was presumed that the outer layer might have peeled off due to the use of the actual product in combination with a metal needle.However, the definite cause of occurrence could not be determined since the actual sample was not available for analysis.The ifu (instructions for use) of this product includes the following statement: "do not manipulate or withdraw the glidewire through a metal entry needle or a metal dilator.Manipulation and/or withdrawal through a metal entry needle or a metal dilator may result in destruction and/or separation of the outer polyurethane coating requiring retrieval.A plastic entry needle is recommended when using this wire for initial placement.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RADIFOCUS GUIDEWIRE OR GLIDEWIRE OR GLIDEWIRE M
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
TERUMO CORPORATION, ASHITAKA
150 maimaigi-cho
fujinomiya city, 418
JA  418
Manufacturer (Section G)
TERUMO CORPORATION, ASHITAKA
reg. no. 9681834
150 maimaigi-cho
fujinomiya city, 418
JA   418
Manufacturer Contact
stephanie handy
reg. no. 2243441
950 elkton blvd
elkton, MD 21921
9499890491
MDR Report Key14192173
Report Number9681834-2022-00074
Device Sequence Number17237770
Product Code DQX
UDI-Device Identifier04987350773654
UDI-Public04987350773654
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K863138
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2023
Device Model NumberN/A
Device Catalogue NumberGS3509
Device Lot Number210914
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/01/2022
Supplement Date Received by Manufacturer04/26/2022
Initial Report FDA Received Date04/25/2022
Supplement Report FDA Received Date04/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SPECTRONETICS 0.035 QUICKCROSS CATHETER.
Patient Age64 YR
Patient SexMale
Patient EthnicityNon Hispanic
Patient RaceBlack Or African American
-
-