Model Number 62-3401R
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 04/08/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Noiles femoral implant broken on the sleeve connection.Doi: (b)(6) 2018; dor: (b)(6) 2022; unknown side.
 
Additional Manufacturer Narrative
Product complaint # : (b)(4).Investigation summary : the device associated with this complaint was received for examination.The photo investigation and visual examination of the returned device confirmed the reported allegation.The device was found broken.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : a manufacturing record evaluation was performed for the finished device , and no non-conformances / manufacturing irregularities were identified.Device history review
=
> a manufacturing record evaluation was performed for the finished device , and no non-conformances / manufacturing irregularities were identified.
 
Event or Problem Description
Additional information received: a.Patient demographics the following are the types of patient demographics we require: relevant comorbidities, date of implantation (b)(6) 2018, date of revision (b)(6) 2022, operative side right, age at the time of the procedure 65, weight 120 kg, height 1,80m, and activity level.Active.B.Excerpts from the clinical correspondence: please provide all correspondence relevant to the reported event for example: onset of symptoms, trauma, other devices, leg length discrepancy, gait patterns, medical investigations or any other issues the complainant feels important.The only thing we know is that the patient fell.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SROM NRHFEM W/PIN MED RT 71X66
Common Device Name
LPS AND S-ROM : KNEE FEMORAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY RAYNHAM, A DIV. OF DEPUY ORTHO 1219655
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key14192363
Report Number1818910-2022-07387
Device Sequence Number17255073
Product Code KRO
UDI-Device Identifier10603295179474
UDI-Public10603295179474
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K896048
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number62-3401R
Device Catalogue Number623401R
Device Lot NumberHF3561
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/05/2022
Supplement Date Received by Manufacturer04/27/2022
06/30/2022
Initial Report FDA Received Date04/25/2022
Supplement Report FDA Received Date05/05/2022
06/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/21/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
LPS UNIV TIB HIN INS MED 12MM; LPS UNIV TIB HIN INS MED 12MM; MBT REVISION CEM TIB TRAY SZ 5; MBT TRAY SLEEVE POR M/L 29MM; SROM NRHFEM W/PIN MED RT 71X66; UNIVERSAL FEM SLV DIS POR 40MM; UNIVERSAL STEM 75X16MM FLUTED; UNIVERSAL STEM 75X16MM FLUTED
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age69 YR
Patient SexMale
Patient Weight120 KG
-
-