C.R. BARD, INC. (BASD) -3006260740 POWERPORT ISP M.R.I. IMPLANTABLE PORT WITH ATTACHABLE 6F CHRONOFLEX; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
| Model Number |
1806061 |
| Medical Device Problem Codes |
Difficult to Flush (1251); Suction Problem (2170)
|
| Health Effect - Clinical Codes |
Pain (1994); Swelling/ Edema (4577)
|
| Date of Event |
08/17/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
As the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiration date: 04/2022).
|
| |
|
Event or Problem Description
|
|
It was reported that sometime post port placement, the device allegedly had no blood return and difficult to flush.It was further reported that the patient experienced pain and swelling during infusion treatment and flushing of the device.Reportedly, the port can no longer be used for treatment due to patients symptoms experienced during use of device.The current patient status is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: (expiry date: 04/2022).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: device not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that sometime post port placement, the device allegedly had no blood return.It was further reported that the patient experienced pain and swelling during infusion treatment and flushing of the device.Reportedly, the port can no longer be used for treatment due to patients symptoms experienced during use of device.The current patient status is unknown.
|
| |
|