| Model Number |
BIO-COMP INTER-FERENCE SCRW |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/31/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during anterior cruciate ligament surgery during the first turns of the screw, when inserting, the screw thread on the screw head twisted off directly.There was no harm for patient, operator or third party.The surgery was finished successfully with a new device with part number ar-5035tc-09.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint confirmed.One unpackaged ar-5028c-08 was received for investigation.Visual inspection identified breakage across the distal-most threads of the returned biocomposite screw.The method used to prepare the bone socket during the procedure, as well as the bone quality encountered, were not provided.A probable cause is attributed to improper bone preparation.
|
| |
|