• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® INLAY OPTIMA® URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® INLAY OPTIMA® URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE Back to Search Results
Model Number 787726
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event or Problem Description
It was reported that there was no issue with the package, but when it was opened, the ureteral stent was found to be ruptured and the guidewire was missing.
 
Event or Problem Description
It was reported that there was no issue with the package, but when it was opened, the ureteral stent was found to be ruptured and the guidewire was missing.
 
Additional Manufacturer Narrative
The reported event was inconclusive because no sample was returned for evaluation.A potential root cause for this event could be, "inappropriate package design".As no patient involvement was reported the device was not used though it is intended for treatment.As the device was not returned for evaluation it was unknown if it had met all relevant specifications or resulted in the reported event.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "avoid improper handling of stent such as bending, kinking, tearing, etc.Misuse could damage the overall integrity of the stent.Care should be exercised when removing the stent from the inner polybag to eliminate tearing or fragmentation." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® INLAY OPTIMA® URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE
Common Device Name
URETERAL STENT
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key14192720
Report Number1018233-2022-02906
Device Sequence Number4406863
Product Code FAD
UDI-Device Identifier10801741015523
UDI-Public(01)10801741015523
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K043193
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/16/2024
Device Model Number787726
Device Catalogue Number787726
Device Lot NumberNGFR2781
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/01/2022
Supplement Date Received by Manufacturer06/15/2022
Initial Report FDA Received Date04/25/2022
Supplement Report FDA Received Date07/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-