C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® ALL-SILICONE TEMPERATURE-SENSING FOLEY CATHETER Back to Search Results
Model Number 119112
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/31/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the water did not come out from the foley catheter balloon during pretest.Per follow-up information received via ibc on (b)(6) 2022, stated that it was difficult to deflate the foley catheter balloon and the damage in the balloon was unknown.
 
Additional Manufacturer Narrative
The reported event was unconfirmed because the reported failure could not be reproduced.The reported failure is within specification as the reported failure could not be reproduced.The product was not used for patient treatment or diagnosis.The product had not caused the reported failure.No root cause could be found because the reported event was unconfirmed.Visual evaluation of the returned temp sensing foley catheter.Visual inspection of the sample noted no obvious visible defects.The catheter balloon was inflated with 5.5 ml methylene blue solution.The catheter was allowed to rest with no observed leaks and deflated in 1 min 38.42 seconds without issue or cuffing.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.As the reported event is unconfirmed a labeling review is not required.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Event or Problem Description
It was reported that the water did not come out from the foley catheter balloon during pretest.Per follow-up information received via ibc on 06apr2022, stated that it was difficult to deflate the foley catheter balloon and the damage in the balloon was unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® ALL-SILICONE TEMPERATURE-SENSING FOLEY CATHETER
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key14193548
Report Number1018233-2022-02922
Device Sequence Number4406895
Product Code EZL
UDI-Device Identifier00801741046018
UDI-Public(01)00801741046018
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K070582
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/24/2025
Device Model Number119112
Device Catalogue Number119112
Device Lot NumberNGFZ1650
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/22/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/01/2022
Supplement Date Received by Manufacturer10/05/2022
Initial Report FDA Received Date04/25/2022
Supplement Report FDA Received Date10/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/06/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-