Model Number 139112EXT
Medical Device Problem Code Delivered as Unsterile Product (1421)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Reported event is confirmed.Customer event ¿insufficient heatseal, (sterile) pouch or blister pack¿ was confirmed based on device evaluation.Received two 139112ext in unopened original packaging.Lot number was verified.Performed a visual inspection of the device, there is an open channel in the seal on one out of the two packages.Performed a functional inspection, the devices were dye leak tested which indicated that the packaging had an insufficient heat seal on the packaging with the channel.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.A two-year review of complaint history revealed there has been a total of (b)(4) reports, regarding (b)(4) devices, for this device family and failure mode.During this same time frame (b)(4) devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be 0.0001.Conmed encourages the inspection of all medical equipment, labeling and packaging prior to use.This issue will continue to be monitored through the complaint system to assure patient safety.
 
Event or Problem Description
During incoming inspection, the distributor rejected this device, 139112ext, for an insufficient heatseal.There was no contact with the patient as this was found during incoming inspection.This will be reported as a malfunction with potential for injury upon reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRCLN 4 IN. BLD ELEC W/ EXTND INSUL
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
CONSOLIDATED MEDICAL EQUIPMENT COMPANY
ave. alejandro dumas 11321
complejo industrial chihuahua
chihuahua 31136
MX  31136
Manufacturer (Section G)
CONSOLIDATED MEDICAL EQUIPMENT COMPANY
ave. alejandro dumas 11321
complejo industrial chihuahua
chihuahua 31136
MX   31136
Manufacturer Contact
robin drum
11311 concept blvd
largo, FL 33773
8653881978
MDR Report Key14193601
Report Number3007305485-2022-00051
Device Sequence Number4556669
Product Code GEI
UDI-Device Identifier40653405034774
UDI-Public(01)40653405034774(17)260128(10)202101294
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K991855
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date01/28/2026
Device Model Number139112EXT
Device Catalogue Number139112EXT
Device Lot Number202101294
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/24/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/30/2022
Initial Report FDA Received Date04/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-