Imaging evaluation was revised after additional images were received.Explant evaluation was completed.Imaging evaluation: there is contrast within the portal venous system.The arterial scan timing is slightly late providing this visualization of the portal venous system.Additional review of the lumbar arteries and pooling in the sac in the delayed phase is consistent with a type 2 endoleak.Proximal aorta: there is contrast in the lumbar artery visualized communicating with the aorta and outside the graft.Proximal aorta: there is no evidence of a contrast pooling distal to the proximal seal zone which would indicate a type 1a endoleak.There is contrast visualized within the lumbar artery communicating with the aneurism sac.The contrast pooling is greater within the delayed phase consistent with a type 2 endoleak.There is no contrast visualized outside the limbs which would indicate a type 1b endoleak.Vs2 in every time point there is a lumbar artery communicating with the aneurism sac with corresponding contrast pooling in the aneurism sac.There is no evidence to conclude anything other than a type 2 endoleak as previously documented.Explant evaluation: the devices were returned to w.L.Gore & associates for investigation.Submitted in formalin were three gore excluder aaa endoprostheses; one trunk ¿ ipsilateral leg endoprosthesis fragment (tf) containing one contralateral leg endoprosthesis fragment (clf) and one contralateral leg endoprosthesis (cl).The device fragments had been transected prior to arrival at w.L.Gore& associates.There was scant, tan to red brown, firmly adherent tissue on the abluminal and luminal surfaces.Multifocal areas of firmly adherent, white tissue were present in the lumen of the tf.The lumens were all widely patent.Histopathologic sampling was not performed, due to the lack of adherent tissue.The specimens were subjected to an enzymatic digestion process to remove biologic debris.Following digestion all specimens were examined for material disruptions with the aid of a stereomicroscope.Certain material disruptions identified (e.G., electrocautery, transections, and forceps marks) were consistent with those caused by interaction with surgical instrumentation (i.E., electrocautery, scalpel/scissors, forceps/clamps), likely used during the explant procedure.The other material disruptions identified (e.G., constraint sleeve partial material separation, radial film disruption, and wire fractures) were likely caused by in vivo wear.There is no evidence that these finding would impact device performance.
|